A Feedback Control Approach to Organic Drug Infusions Using Electrochemical Measurement
Abstract
Goal: Target-controlled infusion of anesthesia is a closed-loop automated drug delivery method with a computer-aided control. Our goal is to design and test an automated drug infusion platform for propofol delivery in total intravenous anesthesia (TIVA) administration. Methods: In the proposed method, a dilution chamber with first-order exponential decay characteristics was used to model the pharmacodynamics decay of a drug. The dilution chamber was connected to a flow system through an electrochemical cell containing an organic film-coated glassy carbon electrode as working electrode. To set up the feedback-controlled delivery platform and optimize its parameters, ferrocene methanol was used as a proxy of the propofol. The output signal of the sensor was connected to a PI controller, which prompted a syringe pump for feedback-controlled drug infusion. Results: The result is a bench-top drug infusion platform to automate the delivery of a propofol based on the measurement of concentration with an organic film-coated voltammetric sensor. Conclusion: To evaluate the performance characteristics of the infusion platform, the propofol concentration in the dilution chamber was monitored with the organic film-coated glassy carbon electrode and the difference between the set and measured concentrations was assessed. The feasibility of measurement-based feedback-controlled propofol delivery is demonstrated and confirmed. Significance: This platform will contribute to high-performance TIVA application of intravenous propofol anesthesia.
Publication Title
IEEE Transactions on Biomedical Engineering
Recommended Citation
Myers, M., Li, Y., Kivlehan, F., Lindner, E., & Chaum, E. (2016). A Feedback Control Approach to Organic Drug Infusions Using Electrochemical Measurement. IEEE Transactions on Biomedical Engineering, 63 (3), 506-511. https://doi.org/10.1109/TBME.2015.2464771