An esophageal probe for measuring three-dimensional electric fields during external cardiac defibrillation

Abstract

External defibrillation is a common treatment for the cardiac arrhythmia atrial fibrillation. Electrode placement has been shown to affect defibrillation efficacy and required energy levels. We suggest investigating the relationship between esophageal electric fields (EEFs) and atrial defibrillation thresholds to determine the feasibility of creating patient-specific electrode placements using EEFs. This study presents the design and implementation of an esophageal probe (EP) that accurately measures three-dimensional electric fields. The root-mean-square error of the EP was 1.69% as determined by measurements performed in an electrolytic tank. The EP also performed well during in vivo testing in a pig. There was a strong positive relationship between EEF 2s and applied energy during defibrillation strength shocks. The EEF measurements were also repeatable, with less than 4.24% difference between repeated shocks. This is the first description of a probe designed specifically for measuring electric fields in the esophagus. © 2012 Institute of Physics and Engineering in Medicine.

Publication Title

Physiological Measurement

Share

COinS